Diagnosing An Error Message With Your Scale:

 

Err 30:

 

Illegal Zero – This might come up when the scale is first turned on, or when the zero button is pressed. (It is most often followed by the scale returning to the original weight reading on the scale.)

Possible Solutions

 

-         Refer to the Calibration Manual. Calibration step “-0- 23” sets the zero range for the scale. This is the amount of weight that will be allowed to be zeroed off and is normally set to the scales full capacity. Any weight greater then the value entered in this step will give the illegal zero error and not allow the weight to be zeroed off.

-         Refer to the Calibration and Programming manual. Calibration step “Cnd 26” is a selection for setting the scale up to meet Canadian Weights and measures specifications. This specification does not allow the scale to zero more than 4% of the scale’s capacity. For U.S. applications, the step should be set to “NO”.

-         If none of the above solutions work, or you are not comfortable with doing the programming on your scale, contact your Pennsylvania Scale distributor for service.

 